2.5.4.1 Safety Test
2.5.4.2 Note for the Administration of "Full Backup" Data
Voluson® S6/S8 Basic User Manual
5433669-100 Revision 4
Scan time limits: According to respective national regulations, and according to the
manufacturer recommendations for the medical-technical system.
Range:
a)
Visual
Housing, connection, operating elements, display facilities,
inspection:
labels, accessories, user manual.
b)
Functional test:
Checking of functions (according to user manual), check also
modular combinations and common operability of system and
accessories.
c)
Electric test:
Checking of the electric safety of system combinations
according to IEC60601-1 or respective national regulations.
For safety reasons, avoid handling fluids in the vicinity of the system.
Item
Safety Test
Console Leakage
Annually
Current Checks
Peripheral Leakage
Annually
Current Checks
Surface Probe
Annually
Leakage Current
Checks
Endocavity Probe
Annually
Leakage Current
Checks
All settings and patient data created since last full backup are NOT backed-up! It is highly
recommended to create a full backup of settings and patient data regularly.
When the Full Backup is stored on a network drive, it may be desirable to move the data (e.g.,
for backup or maintenance).
The directory structure of the full backup data is as follows:
Every "Full Backup" resides in a subfolder of the main "
the drive. For example: Z:\fullbackup.
Also after corrective maintenance or as required
by your facilities QA program.
Also after corrective maintenance or as required
by your facilities QA program.
Also after corrective maintenance or as required
by your facilities QA program.
Also after corrective maintenance or as required
by your facilities QA program.
For more information see 'Backup' on page 13-47.
fullbackup" -folder found at the root of
Safety
Notes
2-13