Entering Standard Volume Data; Entering Settings - GE CTF878 Programming Manual

Clamp-on tag flowmeter
Hide thumbs Also See for CTF878:
Table of Contents

Advertisement

Entering Standard
Volume Data

Entering Settings

Programming Site Data
The Standard Volume option enables you to set base measurements
and specific values for pressure and temperature inputs and to select
the corresponding option cards. To enter this option, scroll to
Standard Volume on the Program Menu and press [
The screen appears similar to Figure 1-27 below. To step through
each parameter, press the [ ] key. Refer to the Program Menu Map,
Figure 1-34 on page 1-37.
Inputs
Settings
Standard Volume
Enabled
Base Pressure
Atmospheric Pressure
Base Temperature
Figure 1-27: Standard Volume Window - Settings
1. The first prompt allows you to enable or disable the Standard
Volume settings.
a. Use the arrow keys to move to the appropriate choice.
b. Press [
ENTER
] to confirm your choice.
Note: If you selected
Enabled
for the three categories. If you selected
be able to access the three boxes or the Inputs entry.
2. To enter numbers for the three categories:
a. Press the [ ] key to move from one box to the next.
b. Press [ENTER] and use the number keys to enter a value for
each category.
c. Press [ENTER] to confirm each entry.
Standard Volume
Disabled
1.01
1.01
60
Cancel
, continue on to step 2 to enter values
Disabled
February 2009
ENTER
].
Bar
Bar
°F
OK
, you will not
1-27