Entering The Standard Volume; Mass Flow Parameters - GE TransPort PT878GC User Manual

Portable gas flowmeter
Table of Contents

Advertisement

Entering the Standard Volume/

Mass Flow Parameters

The Standard Volume/Mass Flow option enables you to correct
standard volumetric units for pressure and temperature, as well as
to disable or enable mass flow. To enter this option, scroll to the
Std Vol/Mass Flow entry on the Program Menu and press
The screen appears similar to Figure 4-7 below. To step through
each parameter, press the [ ] key. Refer to Figure A-5 on
page A-5 of Appendix A, Menu Maps.
Settings
Standard_Volume
Enabled
Atmospheric Pressure
Figure 4-7: The Standard Volume Option Window
1. The first prompt asks if you want to disable or enable the
Standard Volume option.
a. Use the [ ] and [ ] keys to scroll to the appropriate
radio button.
b. Press
[ENTER]
Note: If you select "Disabled," you cannot select any other
prompts in this window.
Programming Site Data
Std Vol/Mass Flow
Mass Flow
Inputs
Base Pressure
Base Temperature
Cancel
to confirm your selection.
Disabled
1.0
Bars
Bars
1.0
°C
15.5
l
April 2009
[ENTER]
.
OK
4-17