Entering Factory Calibration Data - GE XDP-H2 User Manual

Panametrics explosion-proof display package
Table of Contents

Advertisement

January 2008
Configuring the 4-20 mA
Analog Input (cont.)
Entering Factory
Calibration Data
5-16
Now, the XDP analog output is driven to exactly 20 mA.
4-20mA-I..20mA-Trim
xx.xxx
s
xx
t
4-20mA-Input
Test
Trim
To continue programming the setup menu, proceed to the appropriate
section for instructions. Otherwise, press
live measurement mode.
To enter the factory calibration data, access the setup menu as
described on page 5-4 and complete the following steps:
Range
Cal
Setup
Serial Number
4-20mA-Input
Factory Calibration
At the next prompt, select the range to be calibrated (H2/AIR is used
as an example in this manual).
Gas Curve:
H2/AIR
H2/CO2
AIR/CO2
At the next prompt, enter the number of data points available for
calibrating the chosen gas curve.
Gas-Curve: H2/AIR
Enter # of Points
xx
s
Press the [ ] and [ ] keys
until the meter reads exactly
20 mA. When done, press
[ENTER]
.
[ESC]
Press
to return to the
setup submenu.
[ESC]
four times to return to
Use the [ ] and [ ] keys to
[Factory Calibration]
select
[ENTER]
press
.
Use the [ ] and [ ] keys to
[H2/AIR]
select
and press
[ENTER]
.
Use [ ] and [ ] to position the
cursor under the desired char-
Œ
acter. Then, use [
] and [
increment the value. Press
[ENTER]
when done.