GE CTF878 Programming Manual page 30

Clamp-on tag flowmeter
Hide thumbs Also See for CTF878:
Table of Contents

Advertisement

February 2009
Entering Analog Outputs
(cont.)
1-22
Select Measurement
DIAG
Velocity
AuxIn
Pressure
Temperature
Volume
Fwd Totalizer
Rev Totalizer
StdFwd Totalizer
No Unit
Figure 1-23: Data Source Selection Window
4. The next prompt requests the span (full scale) value for the analog
output. This value represents the 20 mA output (in flow units).
a. Press
[ENTER]
to open the text box.
b. Use the numeric keys to enter the desired value.
c. Press
[ENTER]
to confirm your selection.
5. The last prompt, On Error, asks you to select how the CTF878 will
handle the analog outputs in the event of a fault condition. The
meter offers three alternatives:
Hold Last Value (hold the last good reading)
Force Low (force the reading to 0 or 4 mA)
Force High (force the reading to 20 mA).
a. Press
[ENTER]
to open the drop-down menu.
b. Scroll to the desired selection.
c. Press
[ENTER]
to confirm your selection.
You have completed entering data in the Analog Output option.
To confirm the entries and return to Operate mode, press
The CTF878 returns to Operate Mode.
To leave the window without confirming the entries, press
[ESC]
(Cancel) or the
key. The CTF878 returns to Operate Mode.
m/s
Cancel
OK
Programming Site Data
[F3]
(OK).
[F2]