Solar Panel Charging System Test; Table 5-2 Specifications For Solar Panels - ABB XSeries G4 6200 User Manual

Flow computer
Table of Contents

Advertisement

2) Replace the battery with a known good battery using the Battery Pack
3) Reconnect power to the charger regulator. If the battery pack is charged
4) Measure the charging voltage from the solar panel at the charger regulator.
5) If the unit receives charging voltage from the AC charger, perform the AC
Panel
Max
Volts @ PMax
5
4.5W
16.5V
10
10W
16.8V
20
20W
16.8V
30
30W
16.8V

5.3.4 Solar panel charging system test

If the system setup includes a solar panel charger connected to the 6270 Optional
Equipment Unit (OEU) and is not supplying the required voltage and current to the
G4 EX unit, the user may need to test the solar panel.
The following instructions contain the steps required to test the solar panel.
5.3.4.1
-
-
5.3.4.2
1) Measure the solar panel voltage at the charger regulator. Use a DMM
2) Check the solar panel angle and direction. In the northern hemisphere, the
Page 5–10
Installation procedure in Chapter 2.
through an AC charger, move to step 5; otherwise, continue to step 4.
Use a DMM that connects the (+) and (-) leads to the (+) and (-) solar panel
wires. Loaded voltage should be greater than or equal to the specifications
listed in Table 5–2. If the voltage is within range, the battery was bad.
If the loaded voltage is not above the minimum, perform the Solar Panel
Troubleshooting Test found within this chapter.
Charger Troubleshooting Test found within this chapter.
Table 5–2 Specifications for solar panels
Open Circuit
20.5V
21V
21V
21V
Required equipment
Digital Multi-meter with 0-20 VDC range
Required resistors for testing specific panels as listed in Table 5–2
In continuous low sunlight conditions, the unit may not
supply the required voltage. The solar panel should be
positioned so that it receives the most sunlight. Do not place
in a shaded area.
Step-by-step instructions
connecting the (+) and (-) leads to the (+) and (-) solar panel wires. Loaded
voltage should be greater than or equal to the specification listed in Table
5–2. If the solar panel is not above the minimum, replace the solar panel,
and continue to step 2.
panel should face due south. In southern hemispheres, the panel should
face due north.
Load Resistance
Loaded Voltage
50 Ω 25W
13.5V
30 Ω 25W
17.5V
15 Ω 25W
17.8V
10 Ω 40W
17.8V
2104349 rev. AD