Calibration; Required Test Equipment; Table 4-1 Calibration Configurable Parameters - ABB XSeries G4 6200 User Manual

Flow computer
Table of Contents

Advertisement

10) Gain access to the G4 EX board by loosening the countersunk hex socket
11) After the end cap has been removed, gently pull the display overlay and
12) After the graphic overlay plate is off, gently remove the lithium battery
13) Upon removal of the lithium battery connector, snap the board out of the
14) Unplug the raw sensor connector cable (J6) and the ribbon cable (J1) from
15) Gently remove the board from the G4 EX device.
16) Loosen the 2" mounting pole U-bolts. Rotate/lower the unit a sufficient
17) Loosen the hex socket head set screws on the neck of the unit using a 3/32"
18) Replace the transducer and raw sensor board with a new/repaired/spare
19) Reconnect the high/low manifold lines, and restore pressure.
20) Reassemble the unit by performing steps 10-14 in reverse order.
21) Following the instructions in section 4.3, restore the station files.

4.13 Calibration

The calibration mode allows the user to calibrate, check and zero the static and
differential pressure. Additionally, this mode allows the user to set the bias for the
resistance temperature detector (RTD). During the initial calibration, the
parameters shown in Table 4–1 will need to be configured.
Parameter
RTD Installed
RTD Bias
Temperature Base (Tb)
Used Fixed Temperature (TF)

4.13.1 Required test equipment

The following test equipment is required to calibrate the G4 EX EXIMV:
-
-
-
-
-
2104349 rev. AD
locking set screw in the front end cap. Use a 1/16" hex wrench to perform
this task. When completed, unscrew the end cap.
graphic overlay plate away from the snap on standoffs.
connector (J5) from the G4 EX board.
device. Please note that the board is still connected to the device through
the cables in the back.
the G4 EX board.
distance so that the transducer can be removed.
hex wrench. Continue until the transducer turns freely. Gently turn counter-
clockwise until the transducer is free from the enclosure.
unit of the same type and pressure rating.
Table 4–1 Calibration configurable parameters
PCCU32 version 7.17 or later
Deadweight tester or equivalent calibration standard
Test gauge capable of dual range measurement (PSIG and inches)
Barometer or another means which can determine barometric pressure
Nitrogen or compressed air source
Default Value
Units
No
0.00
Degrees F
60.000
Degrees F
No
Page 4–19