Testing The Patient Module/Toco+ Transducer: Mecg Mode - Philips Avalon FM20 Service Manual

Hide thumbs Also See for Avalon FM20:
Table of Contents

Advertisement

7 Testing and Maintenance
CAUTION
The tip of the spiral electrode is sharp. Take care not to injure your fingers.
or
Cut off the plastic tip of the fetal scalp electrode (containing the spiral and reference electrodes)
from the end of the wires. Strip the insulation from the end of the wires, and connect them to a
patient simulator.
NOTE
We do not recommend the use of a specific patient simulator. The use of a patient simulator does not
allow checking the specification of the ECG-functionality; it allows only a check of general function.
Result: the
Viewing the ECG wave: when configured, you can view the DECG wave on the screen, and any noise
will be visible as additional verification of the effectiveness of the test.
If the test results are not as outlined above, repeat the test with another ECG transducer. If this does
not solve the problem, try the following:
Check all connections.
If the
the adapter cable.
If the problem persists, replace the transducer.
Testing the Patient Module (M2738A)/Toco+ Transducer (M2735A): MECG Mode
Switch on the monitor and the recorder.
1
Connect the patient module or Toco
2
Attach the MECG adapter cable M1363A to the red color-coded socket on the patient module or
3
Toco
94
x
INOP should disappear.
DFHR
Leads Off
x
INOP is still displayed, the DECG adapter cable may be defective. Replace
DFHR
Leads Off
+
transducer
+
transducer to the fetal monitor.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Avalon fm30

Table of Contents