10 Repair and Disassembly
Remove the two screws (one on each side) fastening the RFI bracket to the recorder chassis.
8
9
Partially unscrew the screw nearest the recorder adapter board holding the stepper motor to the
recorder chassis, until the thread no longer protrudes. This allows the necessary clearance for
removing the RFI bracket.
166