General Test Setup for the Performance of Safety Tests

Connect the detachable power cord of the device under test to the safety analyzer's test mains port.
For testing the applied part leakage current, connect all applied parts to the safety analyzer using the
appropriate patient lead or adapter cable. If necessary, repeat the safety test procedure until all available
applied parts have been tested. Refer to the documentation that accompanies the safety analyzer for
further details on how to set up and perform the test. Power on the device under test for all tests.
Protective Earth Resistance Test - Setup Example
Equipment Leakage Current Test - Setup Example
7 Testing and Maintenance
63