12 Specifications
SpO
Alarm Specifications Range
2
Bradycardia

Tympanic Temperature

Complies with:
EN 12470-5 (Clinical thermometers - Part 5:2003: Performance of infra-red thermometers)
ASTM E1965-98 (Infrared Thermometers for Intermittent Determination of Patient
Temperature)
with minor exceptions as noted below.
The fetal monitor additionally complies with ISO 80601-2-56:2009 / EN ISO 80601-2-56:2012
Performance Specifications
Temperature Resolution
Response Time
Temperature Calibrated Accuracy Specifications (out of the Factory)
Ambient Temperature
25.0°C (77.0°F)
16.0°C—33.0°C (60.8°F—
91.4°F)
Temperature Calibrated Accuracy Specifications (after recalibration using Genius 2 Checker/Calibrator)
Ambient Temperature
16.0°C—33.0°C (60.8°F—
91.4°F)
16.0°C—33.0°C (60.8°F—
91.4°F)
ASTM laboratory requirement for IR thermometers in the display range 37.0°C—39.0°C (98.0°F—102.0°F) is
±0.2°C (±0.4°F), whereas for mercury-in-glass and electronic thermometers, the requirement per ASTM standards
E667-86 and E1112-86 is ±0.1°C (±0.2°F).
Clinical accuracy characteristics and procedures are available from Covidien llc on request. To verify the accuracy, use
a certified black body as specified in EN ISO 80601-2-56, Annex C, or use a Genius 2 Checker/Calibrator - available
from Covidien llc under part number 303097.
Clinical repeatability: meets section A.5 of EN ISO 80601-2-56(E) per Covidien llc technical report. Data is available
from Covidien llc on request.
Displayed Temperature Measurement Range
Mode
Ear
Oral (ear + 0.6°C)
Core (ear + 1.04°C)
236
Difference to low limit 0—50 bpm 5 bpm steps
Clamping at 30—100 bpm
Target Temperature
37.7°C—38.9°C (98.4°F—102.0°F)
33.0°C—42.0°C (91.4°F—107.6°F)
Target Temperature
36.0°C—39.0°C (96.8°F—102.2°F)
<36.0°C or >39.0°C
(<96.8°F or >102.2°F)
Range °C
33.0—42.0
33.6—42.0
34.0—42.0
Adjustment
5 bpm steps
0.1°C or 0.1°F
less than 2 seconds
Accuracy
±0.1°C (±0.2°F)
±0.2°C (±0.4°F)
Accuracy
±0.2°C (±0.4°F)
±0.3°C (±0.5°F)
Range °F
91.4—107.6
92.5—107.6
93.2—107.6
Delay
max. 14 seconds

Hide quick links: