11 Upgrades
Option B72 (FM30 only)
1
Remove the top cover assembly as described "Removing the Top Cover Assembly" on page 141.
Remove the noninvasive blood pressure assembly.
2
Remove the front end assembly -- see "Removing the Front End Assembly" on page 200.
3
Remove the connector block frame from the front end assembly (see "Disassembling the Front
4
End Assembly" on page 202). Leave the fetal sensor socket connector block in place.
Remove the blood pressure connector from the bus master board.
5
Join the new SpO
6
218
connector to the noninvasive blood pressure connector.
2