Philips Avalon FM20 Service Manual page 57

Hide thumbs Also See for Avalon FM20:
Table of Contents

Advertisement

Service Event
Tympanic Temperature performance testing
Other Regular Tests and Tasks
Service Event
Visual Inspection
Transducer and Patient Module Testing
Recorder Maintenance
Repairs
Service Event
Repairs when the monitor has been damaged by impact,
liquid ingression, fire, short circuit, or electrical surge
Repairs where the power supply, the mains socket or an
interface board is removed, or replaced, or the protective
earth ground connection is disrupted
Repairs where the NBP pump has been replaced
All repair events involving any of the following:
Noninvasive Blood Pressure connector, SpO
board or the red fetal sensor sockets
Upgrades
Service Event
Software upgrades
Hardware upgrades
Installation of interfaces or hardware where the power
supply, or parameter boards need to be removed
Noninvasive Blood Pressure hardware upgrade M2702AU
Option B71
SpO
hardware upgrade M2703AU Option B72
2
Noninvasive Blood Pressure/SpO
M2703AU Option B73
Test Block(s) Required - Complete these tests
Tympanic Temperature calibration
Test Block(s) Required - Complete these tests
Perform Visual Inspection
See
Regular cleaning and maintenance (see "Transducers and
Patient Modules; Functional Tests" on page 91)
Perform the recorder selftest (see "Fetal Recorder Selftest
Report" on page 98)
Test Block(s) Required - Complete these tests
Perform Visual Inspection, Power On, all Safety tests, and
Full Performance Assurance tests
Perform Visual Inspection, Power On, all Safety Tests,
and Basic Performance Assurance tests
Perform Visual Inspection, Power On Test, all Safety tests,
Basic Performance Assurance test, and NBP Performance
test and calibration
Perform Visual Inspection, Power On, Full Performance
connector/
tests, and all Safety tests
2
Test Block(s) Required - Complete these tests
Perform Visual Inspection, Power On test, and Basic
Performance Assurance test unless otherwise specified in
the Upgrade Installation Notes shipped with the upgrade
Perform Visual Inspection, Power On test, and Basic
Performance Assurance test unless otherwise specified in
the Upgrade Installation Notes shipped with the upgrade
Perform Visual Inspection, Power On test, Basic
Performance tests, and all Safety tests
Perform Visual Inspection, Power On, Noninvasive Blood
Pressure performance, and all Safety tests
Perform Visual Inspection, Power On, SpO
performance, and all Safety tests
hardware upgrade
Perform Visual Inspection, Power On, Noninvasive Blood
2
Pressure/SpO
7 Testing and Maintenance
performance and all Safety tests
2
2
57

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Avalon fm30

Table of Contents