1 Introduction

Repair Strategy

The Service Support Tool software helps you to determine whether a fault is a hardware or software
problem. The main replaceable parts are:
unit exchange for the transducers
replacement of
the top cover assembly
the bottom housing
the power supply assembly
the display assembly
the recorder adapter board
the paper drawer assembly
the paper sensor assembly
the stepper motor assembly
the SpO
the noninvasive blood pressure assembly
the main CPU board
the bus master board
the socket connector block
the transducer cable
the transducer belt button
See "Parts" on page 125 for part numbers, and "Repair and Disassembly" on page 139 for repair
details.
Repair or replacement of individual components on the boards is not supported, and should never be
attempted.
For tests that you are required to perform after repairs, refer to "When to Perform Tests" on page 56.
WARNING
Do not maintain or repair the device in patient vicinity.

Manufacturer's Information

© Copyright 2003 - 2014. Koninklijke Philips Electronics N.V.
All Rights Reserved.
Philips Medizin Systeme Böblingen GmbH
Hewlett-Packard-Str. 2
71034 Böblingen, Germany