Many events that occur during start-up or regular monitoring are logged in the Status Log. The Status
Log can be cleared. Not all entries in the Status Log are errors. You can print the Status Log via the
Support Tool.
The
Status Log
To enter the
Status Log
Product
Appl SW
Config
Boot
Language
Settings
OB
FetRec
(optional)
NBP
(optional)
SpO₂
List of plugged parameters
Select
Status Log
The first column in the log identifies the event class ("C": caused a cold start, "H": caused a hot start,
"N": no restart, for information only). Column 3 and 4 identify the event source and event code.
Column 4 counts the number of occurrences of the event. The last column shows the time and date of
the last occurrence of the event.
The following pop-up keys overlay the SmartKeys:
Clear Stat Log
This key clears the currently displayed Status Log
M2703A
This key switches to the Monitor Revision Window
If an event occurs repeatedly, contact your Philips Service representative.