Preparing for Set Up

3-4-1

Verify Customer Order

1.) After unpacking the equipment, it is important to verify that all items ordered by the customer have
been received. Compare all items listed on the packing slip (delivery note) with those received.
NOTICE
It is recommended to keep and store the shipping carton and all other packing materials (including the
support foams, anti-static plastic cover, etc.), in case the system has to be moved to a different location.
!! NOTICE:
Unpack the devices such a way that packaging can be reused.
For warranty purposes, storage of the above is required for one year from date of purchase.
2.) Visually inspect the system components using the following checklist.
Table 3-5
Step
Item
b
1
Rating Plate
2
Console
3
Control Console
4
Probes
5
LCD Display
6
Fans
7
Rear Panel
8
Covers
9
Peripherals
10
Power Cord
System Voltage
11
setting
NOTE:
Report any items that are missing, back-ordered, or damaged, to your GE Healthcare sales
representative. The contact address is shown in
7
DRAFT (A
EVISION
Damage Inspection Checklist - Voluson E8 / Voluson E6 System
Enter Serial Number: _________________________ (printed on rating plate, see:
Verify that the system is switched OFF and unplugged. Clean the console and control panel.
Physically inspect the control console for missing or damaged items.
After switching on the system, verify the proper illumination of all the control panel buttons.
Check all probes for wear and tear on the lens, cable, and connector. Look for bent or damaged pins on the
connector and in the connector socket on the system. Verify that the EMI fingers around the probe connector
socket housing are intact. Check the probe locking mechanism and probe switch.
Clean the LCD display by gently wiping with a dry, soft, lint-free non-abrasive folded cloth. Inspect the monitor
for scratches and raster burn.
Verify that the system's cooling fans and peripheral fans are operating.
Check the rear panel connectors for bent pins, loose connections and loose or missing hardware. Screw all
the cable connectors tightly to the connector sockets on the panel. Verify that the labeling is in good condition.
Check that all screws are tightly secured in place, that there are no dents or scratches and that no internal
parts are exposed.
Check and clean the peripherals in accordance with the manufacturer's directions.
To prevent EMI or system overheating, dress the peripheral cables inside the peripheral cover.
Check the power cord for cuts, loose hardware, tire marks, exposed insulation, or any deterioration.
Verify continuity. Replace the power cord, as required.
Verify that the Voluson E8 / Voluson E6 ultrasound system is set to the correct voltage.
see:
Section 3-4-2 "System Voltage Settings" on page 3-8
Chapter 3 - Setup Instructions
23, 2012)
UGUST
Recommended Procedure
Contact Information on page
V
E8 / V
OLUSON
OLUSON
S
M
ERVICE
ANUAL
Figure 1-2 on page
1-14)
1-17.
3-7
E6

Advertisement

Table of Contents
loading

Table of Contents