GE H
RAFT
EALTHCARE
D
KTD102576, R
IRECTION
7-8-1
System does not boot up
7-22
7
DRAFT (A
EVISION
System doesn't boot
Is Standby Switch
illuminated ?
Is Standby Switch
orange ?
Press Standby Switch till it
changes to green
System starts booting
Is there any picture on
the monitor ?
Is there any picture on
the monitor ?
Is there any picture on
the touch screen ?
Other error. Please
contact the service
engineer
Figure 7-25 System does not Power On / Boot Up
Section 7-8 - Troubleshooting Trees, Instructions and Tech Tips
23, 2012)
UGUST
Verify that
- Power Cord is connected to Mains
- Power Switch is turned on
(on the backside of the system)
Check cables between
Console and Monitor
Press Standby Switch
for approximately 5 sec.
until system is down.
Shut off the power switch
Wait for 10 sec.
Shut on the power switch again
V
E8 / V
OLUSON
OLUSON
S
M
ERVICE
ANUAL
E6