Hardware Configuration Detection - GE Voluson E8 Service Manual

Hide thumbs Also See for Voluson E8:
Table of Contents

Advertisement

GE H
RAFT
EALTHCARE
D
KTD102576, R
IRECTION
7-2-1

Hardware Configuration Detection

1.) Disconnect all probes.
2.) Press the UTILITIES key on the control panel.
3.) In the "Utilities" menu touch SYSTEM SETUP to invoke the setup desktop on the screen.
4.) On the left side of the screen select ADMINISTRATION and then click the
5.) Disconnect all probes.
6.) Click the HW CONFIGURATION DETECTION button (see:
of the page to start board detection.
The system is restarted automatically after the check and the board versions are displayed in the
System Info page (see:
NOTE:
If a missmatch was found the affected boards are listed.
a.) Click the OK button to shut down the system.
b.) Call your Service representative to replace not matching boards.
When starting the system again a new „board detection" is performed automatically.