ACM manual control
This mode displays the ACM 150 monitor flow schematic. It is the same schematic
displayed in automatic mode,
valve.
When you enter this mode, all valves are inactive. Turn a valve on by
selecting the blue circle area.
2-way valves show with green circle fill when they are selected and activated
to open. Flow can pass through these open valves. Select the blue square
again to close the valve.
 
 
Figure 74. Manual Control - option: Venturi Supply with Line Leak
In the Manual Mode you also have control over the venturi pump and the secondary
pump. The venturi pump requires an air supply, which is off in the default mode.
Select S9 to open the air supply. Alternately, you can use the secondary pump as a
vacuum source in this test mode. The booster pump runs automatically in this mode,
it is not subject to manual control.
 
View valve
cycles, except you have control over each