ACM 150
Figure 43. Trend graph display
Archive view
The ACM 150 monitor stores the infrared spectrum data from all its modes for the
past 30 days. The Archived View allows you to selectively review and identify these
spectra.
When you select Archive View, the menu shown in
keys on the keyboard to enter the desired date and the time interval. Next, choose
any of the three Scan Types, and then the desired Status Results. The following rules
apply to the Scan Types:
Normal: Scans collected in the normal scan mode. Generally, only the Alarm 1 and
Alarm 2 normal scans are important to the user. Validate some of these Alarm 1/
Alarm 2 incidents by reviewing the spectra (or emailing spectra to Honeywell
Analytics for review and comment). If you press "Load Data", the data field
44)
red. Narrow the time to a
one-hour period to hide spectra when no incidents occurred.
Figure 44. Archived view
Demand and Lock-On: Scans collected in either the Demand or Lock-On modes.
 
Figure 44
display. Use the number
75
(Figure