Honeywell ACM 150 Manual To Installation, Operation, And Maintenance page 136

Air composition monitor
Table of Contents

Advertisement

 
ACM 150
Resolving problems
Table 6
lists service messages that might be generated by the ACM 150 monitor
software. Troubleshoot the problems by following the recommended actions. Use
'refer to' for guidance, additional information and/or step-by-step procedures.
Table 6. Service request messages
Service Message
Possible line leak prob-
lems in line xx
Possible line leak prob-
lems in lines (all)
Energy (ZPD) value too
low (or too high)
No data from bench
System stopped
System started
No active scans are
defined
Cannot establish com-
munication with control
boards
Cannot update relays
Cannot load factory
background spectra
Energy (ZPD) address
out of range or can't
determine ZPD
Scan cancelled
Incorrect data length
(short/long scan)
Can not evaluate equa-
tion for relay XX
 
Recommended Action
Test the line and its
check valve for leaks
Check venturi regulator
& air supply Check the
line leak setup
Evaluate the optical
path
Check the bench for a
malfunction Evaluate
optical path
No action required
No action required
Set up an Order of Scan •
Verify that the cables
and power cords are
connected to the con-
trol boards
Verify that the cables
and power cords are
connected to the relay
board(s)
Check for missing file
or path error (should not
occur after initial start-
up)
Too little reaching de-
tector
Normal operation
Poor connection at FTIR
data cable
SQL error in relay setup,
advanced tab
136
Refer To
Line leak tests
(optional)
CDA panel
Line leak test setup
procedure
Evaluating the FTIR
optical path
Evaluating the FTIR
optical path
Stopped at user request
Stopped at user request
Order of scan
Interconnecting
cables
Interconnecting
cables
Contact Honeywell
Analytics
Evaluating the FTIR
optical path
 
Evaluating the FTIR
optical path
Virtual relay setup.