ACM 150
Resolving problems
Table 6
lists service messages that might be generated by the ACM 150 monitor
software. Troubleshoot the problems by following the recommended actions. Use
'refer to' for guidance, additional information and/or step-by-step procedures.
Table 6. Service request messages
Service Message
Possible line leak prob-
lems in line xx
Possible line leak prob-
lems in lines (all)
Energy (ZPD) value too
low (or too high)
No data from bench
System stopped
System started
No active scans are
defined
Cannot establish com-
munication with control
boards
Cannot update relays
Cannot load factory
background spectra
Energy (ZPD) address
out of range or can't
determine ZPD
Scan cancelled
Incorrect data length
(short/long scan)
Can not evaluate equa-
tion for relay XX
Recommended Action
Test the line and its
check valve for leaks
Check venturi regulator
& air supply Check the
line leak setup
Evaluate the optical
path
Check the bench for a
malfunction Evaluate
optical path
No action required
No action required
Set up an Order of Scan •
Verify that the cables
and power cords are
connected to the con-
trol boards
Verify that the cables
and power cords are
connected to the relay
board(s)
Check for missing file
or path error (should not
occur after initial start-
up)
Too little reaching de-
tector
Normal operation
Poor connection at FTIR
data cable
SQL error in relay setup,
advanced tab
136
Refer To
•
Line leak tests
(optional)
•
CDA panel
•
Line leak test setup
procedure
•
Evaluating the FTIR
optical path
•
Evaluating the FTIR
optical path
Stopped at user request
Stopped at user request
Order of scan
•
Interconnecting
cables
•
Interconnecting
cables
•
Contact Honeywell
Analytics
•
Evaluating the FTIR
optical path
•
Evaluating the FTIR
optical path
•
Virtual relay setup.