Air composition monitor
Table of Contents

Advertisement

 
ACM 150
Kinked tube testing
Allows the detection of kinked sample tubes by analyzing the Sample Flow
pressure. If a tube is kinked or bent severely and the pressure does not reach the
configurable setpoint KinkedTubeThreshold, a service event is generated
indicating the port and the pressure value. The Kinked Tube Test may be triggered
automatically or manually. For automatic triggering, the time of day and the interval
may be chosen. For triggering the Kinked Tube Test manually, the user needs to
login at the console and press the button "Perform an Entire Kinked Tube Test..."
as shown below.
Figure 69. Hardware servicing
Software Servicing
Database Statistics
The screen "Software Servicing" may be used to analyze and control the database
on the ACM 150 monitor and is designed for our Field Service Engineers.
Figure 70. Software Servicing
Clicking "Get Database Statistics" will load a list of key parameters describing the
database size and composition. These parameters serve no purpose for the ordinary
user and are therefore not explained here.
 