Nibp Calibration - GE eBike II Series Service Manual

Table of Contents

Advertisement

nIBp C
alIBratIon
This function allows the calibration of the internal pressure
measurement system.
Select „NIBP Calibration" in the service menu and confirm
with ENTER (or SELECT).
Offset calibration
During this calibration step NO pressure should be present
in the system.
The cuff connector (self closing) has to be opened (use cuff
connector).
Remove cuff
Connect the cuff connector with a short tube
The actual pressure value is displayed.
Press SET (or START)
The zero value is stored
Calibration with 200 mmHg
Connect the complete measurement system (hand pump,
manometer and compensating container) as described
under „NIBP test".
Using the hand pump, apply a fixed pressure of
200 mmHg to the system
(control by the external manometer)
Wait some seconds until the pressure is stable -
correct the pressure if necessary
Press SET (or START)
The NIBP calibration is finished
2018112-069 Rev. F

NIBP Calibration

Offset:
p
reSSure calibration
NIBP Calibration
Apply
p
reSSure calibration
eBike II, eBike II L, eBike II EL
0
mmHg
SET
-
offSet
200
mmHg
SET
-
gain
- 73 -