tatIC CalIBratIon wItH CalIBratIon weIgHt
The alignment and calibration procedures (static load
calibration, dynamiv load calibration, NIBP calibration) are
identical for all eBike II ergometers.
All calibration processes are started in the service menu.
To activate the service menu the ergometer has to be
switched off and on again.
As soon as the self test appears on the ergometer display,
press the two softkeys
synchronously and hold:
To prevent settings from being changed by chance, the
service menu is „protected" by a password.
Select „999" with the arrow keys and confirm with SET.
The service menu appears.
Select „„Static load calibration" with the arrow keys and
confirm with ENTER.
Hint
To perform a static load calibration, a 8 kg calibration
weight is needed.
- 66 -
S
ervice Menu control terMinal
p
aSSword for Service Menu
Service Menu
Baudrate 8N1
Static Load Calib.
Analog Comparison
ECG-Startpulse
NIBP Parameter
S
ervice Menu control terMinal
eBike II, eBike II L, eBike II EL
GE
Selftest running
Service Code
999
SET
ENTER