nIBp p
araMeter
If the ergometer load is controlled by the internal
programs, this parameter defines the start of the NIBP
module.
Leadtime
Defines the leadtime of the bloodpressure measurement to
the next load step
Setting range:
40 to 90 sec. in 1 sec. steps
Resting NIBP
Defines whether or not a resting bloodpressure
measurement is taken before starting the stress test.
Setting range:
OFF - ON
pC-M
ode
PC-Mode „Automatic"
If „YES" is activated, the ergometer automatically switches
to remote mode (PC mode) as soon as a valid signal is
received from the interface.
s
C
addle
alIBratIon
(eBike II comfort only)
This parameter is used to adjust the electrical saddle
motor. (display scale 1-40)
Select „TOP Position" and confirm with „ENTER".
By use of the arrow keys, move the saddle to the desired
top position (standard 36 cm above saddle guidance) and
confirm with „SET".
Select „Bottom Position" and confirm with „ENTER".
By use of the arrow keys, move the saddle to the desired
bottom position (standard 7 cm above saddle guidance)
and confirm with „SET".
2018112-069 Rev. F

NIBP Parameter

Leadtime
Resting NIBP
c
onfiguration of
PC Mode
a
utoMatic

Saddle Calibration

TOP
BOTTOM Position
Saddle Calibration
TOP
BOTTOM Position
c
alibration of electrical Saddle adJuStMent
eBike II, eBike II L, eBike II EL
ENTER
nibp p
araMeterS
Automatic
Yes
No
ENTER
M
ode
Position
ENTER
Position
SET
60 sec
ON