This function is used to allign the analog remote signals
(input and output) to the corresponding ECG devices.
The eBike II ergometers have to be equipped with the
optional „COM-Modul". This module is placed onto the
9-pin DSUB connector and screwed to the chassis.
Additionally, a jumper connector has to be placed on the
BKE board.
The COM module handles one analog input and 6 analog
output signals from/to external devices.
Select the analog signal to be aligned and confirm
with ENTER.
- 54 -
(
)
CalIBratIon
eBike II, eBike II L, eBike II EL
Hint
To allign the analog load signals, the corresponding ECG
recorder should be connected as a reference voltage
source.
Disconnect a RS232- or USB cable, if connected.
The allignment of the analog signals can be controlled
after leaving the service menu (e.g. by starting a remote
controlled stress test).
S
ignal aSSignMent of analog connector
Pin 1 :
INPUT Load
Pin 2:
reserved - do not connect
Pin 3:
reserved - do not connect
Pin 4:
OUTPUT BP heartrate
Pin 5:
OUTPUT Load
Pin 6:
OUTPUT Diastole
Pin 7:
OUTPUT rpm
Pin 8:
OUTPUT Systole
Housing:
Signal GND
Analog Calibration
Input
1
Output
1
Output
2
Output
3
Output
4
RETURN
a
lignMent of analog input
(coM
)
Module
Load
Load
Heart.
Syst.
Diast.
/
output SignalS