Alignment „Input: Analog Loadcontrol - GE eBike II Series Service Manual

Table of Contents

Advertisement

a
„I
: a
lIgnMent
nput
IN-Volt:
displays the voltage at the analog input
Set the ECG recorder output to a low load value
(e.g. 20 Watt).
The according voltage is displayed (IN-Volt).
Wait until the voltage has settled.
Select „OFFSET" and confirm with ENTER.
By using the arrow keys set the load value to the ECG
recorder vaue (e.g. 20 Watt) and confirm with SET.
The setting is stored in the ergometer.
Set the ECG recorder output to a high load value
(e.g.300 Watt).
The according voltage is displayed (IN-Volt).
Wait until the voltage has settled.
Select „GAIN" and confirm with ENTER.
By using the arrow keys set the load value to the ECG
recorder vaue (e.g. 300 Watt) and confirm with SET.
The setting is stored in the ergometer.
Setting range:
OFFSET: range from 20 to 1000 Watt in 10 Watt steps
GAIN:
range from 20 to 1000 Watt in 10 Watt steps
2018112-069 Rev. F
l
"
nalog
oadControl
eBike II, eBike II L, eBike II EL
INPUT 1
Load
IN-Volt:
OFFSET
GAIN
ENTER
a
lignMent of analog input
load
INPUT 1
Load
IN-Volt:
OFFSET
GAIN
SET
o
a
ffSet
lignMent of analog input
INPUT 1
Load
IN-Volt:
OFFSET
GAIN
ENTER
a
lignMent of analog input
load
INPUT 1
Load
IN-Volt:
OFFSET
GAIN
SET
g
ain alignMent of analog input
0.10 V
20 Watt
300 Watt
"
0.10 V
20 Watt
300 Watt
"
load
3.05 V
20 Watt
300 Watt
"
3.05 V
20 Watt
300 Watt
"
load
- 55 -