Ergometer Type; System Config - GE eBike II Series Service Manual

Table of Contents

Advertisement

Hint
Do not move the saddle over the „TOP position" (36 cm)
and under the „BOTTOM position" (7 cm - see pictures).
If the motor is moved to the mechanical stop positions,
the eBike power supply will switch off due to overload.
e
t
rgoMeter
ype
This setting appears only in eBike II without NIBP option!
If in PC mode an „I" command is received, the ergometer
sends an identification data string to the PC. / ECG
recorder.
To maintain compatibility to ergoline ergometers of the
former generation, the eBike II ergometer identify as
Ergometrics 900 ergometers.
The identification may be changed by use of this
parameter:
setting ER900 identifies with:
setting ER800 identifies with:
s
C
.
ysteM
onfIg
The hardware configuration of the ergometer is defined in
this setup.
Select the module / option and change the setting to I
if the module is integrated in the actual ergometer.
60 Hz Power:
This setting depends on the mains frequency used and sets
internal filters on the LRE board:
Select 0 for a 50 Hz mains frequency
Select I
for a 60 Hz mains frequency
Hint
Wrong setting of the mains frequency has no influence on the
load regulation or the correct function of the ergometer.
- 52 -
0900P10V243
V3.3
eBike II, eBike II L, eBike II EL
36 cm
c
orrect poSitionS for Saddle adJuStMent

Ergometer Type

ER 900
ER 800
ENTER
c
onfiguration of identification data String
Hint
Must be set to ER900 always !

System Config.

Saddle engine
NIBP-Modul
Analog Interface
60 Hz Power
ENTER
S
ettting of ergoMeter configuartion
7 cm
0 I
0 I
0 I
0 I
2018112-069 Rev. F