Alignment „Ouput: Analog Diastole; Alignment „Ouput: Rpm; Alignment „Output: Ecg - GE eBike II Series Service Manual

Table of Contents

Advertisement

a
„o
: a
lIgnMent
uput
The output voltage representing the diastolic value is
defined with this setting.
It is possible to adjust the voltage and / or the diastolic
value.
The setting is performed as described above under
Alignment „Ouput: Analog Load value"
Setting range:
GAIN (voltage):
GAIN (diastole):
a
„o
: rpM"
lIgnMent
uput
The output voltage representing the actual rpm value is
defined with this setting.
It is possible to adjust the voltage and / or the rpm value.
The setting is performed as described above under
Alignment „Ouput: Analog Load value"
Setting range:
GAIN (voltage):
GAIN (rpm):
a
„o
: eCg"
lIgnMent
utput
(For eBike II REHA Ergometer with integrated ECG amplifier
only)
The zeroline and the amplitude of the analog ECG output
signal is defined.
The setting is performed as described above under
Alignment „Ouput: Analog Load value"
Setting range:
ZEROLINE:
200 to 10000 mV
in 10 mV steps
AMPLITUDE
200 to 10000 mV
in 10 mV steps
2018112-069 Rev. F
d
"
nalog
Iastole
500 to 2000 mV in
2 mV steps
50 to 250 mmHg in
5 mmHg steps
200 to 1000 mV in
2 mV steps
30 to 130 rpm in
5 rpm steps
eBike II, eBike II L, eBike II EL
OUTPUT 4 Diastole
GAIN
GAIN
ENTER
OUTPUT 5
rpm
GAIN
GAIN
ENTER
OUTPUT 6
EKG
ZEROLINE
AMPLITUDE
ENTER
1000 mV
100 mmHg
1000 mV
50 rpm
5000 mV
5000 mV
- 57 -