s
l
C
tatIC
oad
alIBratIon
The static calibration of the load unit is described in the
separate chapter „Calibration / alignment".
d
l
C
.
ynaMIC
oad
al
To perform a dynamic load calibration, a calibrated
dynamic test bench is needed.
The dynamic calibration of the load unit is described in the
separate chapter „Calibration / alignment".
s
u
oftware
pdate
For a description of a software update of the ergometer
(firmware) see the separate chapter „Software update".
B
8n2
audrate
The baudrate can be selected between 1.200 Baud and
115.000 Baud.
Default value for standard ergometers is 4.800 Baud.
Hint
•
The baudrate must be set to 4800 baud always!
Select the desired parameter by use of the right and left
softkeys (↑ ↓) and confirm with ENTER.
eCg s
tartpulse
To configurate the remote start pulse for external ECG
recorders (ergometer is controlled by internal programs).
Select the desired parameter by use of the right and left
softkeys (↑ ↓) and confirm with ENTER.
Adjust the appropriate valuewith the arrow keys and store
the value with ENTER.
Leadtime
Remote pulse leadtime before the next load step
Setting range:
1 to 30 sec. in 1 sec. steps
Pulsewidth
Remote pulse pulse width
Setting range:
100 to 1000 ms in 100 msec steps
- 50 -
Baudrate
S
election of Serial baudrate
ECG - Startpulse
Leadtime
Pulsewidth
c
onfiguration of
eBike II, eBike II L, eBike II EL
8N2
1.200
2.400
4.800
9.600
19.200
ENTER
100 msec
ENTER
ecg
reMote pulSe
2018112-069 Rev. F
10 sec