Electromagnetic flowmeter converter
Table of Contents

Advertisement

Fixed pulse output value
The pulse frequency is set, which is used to output in the fixed output mode through the DO
terminal where the pulse output function is set.
The fixed pulse output value can be set within the range of 0 to 1000pps in increments of 1pps.
Note: When the pulse output is sent out from DO2 terminal, up to 100pps can be output.
Checking the fixed output function setting
Proceed as follows to check the fixed output function.
How to check the fixed value out On/Off setting is shown here.
To check other settings, select the following functions:
Fixed current output value
Fixed pulse output value
Switch operation
Display example
N1: FIXED OUT
N1: FIXED OUT
Select N1: FIXED OUT from the setting item selection menu.
The currently set fixed value output On/Off setting appears.
1:ON
Returns to the menu of setting item selection.
- 115 -
N2: FIX CURR
N3: FIX PULSE
Description