Toshiba L5232 Instruction Manual page 101

Electromagnetic flowmeter converter
Table of Contents

Advertisement

Changing the preset output function
The following is an example to show how to change the preset output function from Output state
hold (1/HOLD) to One-shot pulse output (2:50ms PULSE).
Switch operation
Display example
I2: PRESET FNC
I2: PRESET FNC
I2: PRESET FNC
I2: PRESET FNC
I2: PRESET FNC
Select I2:PRESET FNC from the setting item selection menu.
The currently set preset count output function (HOLD in this
1:HOLD
example) appears.
The cursor appears.
1:HOLD
Press
2:50ms PULSE
The display blinks and the set value is displayed for
confirmation
Press
2:50ms PULSE
disappears and the display returns to the condition that the
setting value is displayed.
Returns to the menu of setting item selection.
- 100 -
Description
to select the setting code 1:50ms PULSE.
again to write the data. Then the cursor
MENU/ENT
6F8A0917