Programming Error
Handling
4-8
1. To program the MMS-3 for error handling, click on the "Errors"
tab.The screen appears similar to Figure 4-6 below.
2. Enter the desired values for the low and high error settings in the
appropriate text boxes.
3. To program the meter's response to errors, pull down the "On Low
Error" list box, and click on one of five choices:
Ignore error
Go to high limit
Go to low limit
Hold last value
Go to error value.
4. Repeat this selection with the "On High Error" list box.
5. Click on
[Apply]
to apply the changes immediately, and on
confirm the changes and close the window. To return the window
to its previous settings, click on the
Figure 4-6: Error Handling Tab
[OK]
[Cancel]
button.
Programming with PanaView