Entering Oxygen Curve Data - GE Moisture monitor series 3 Programming Manual

Panametrics hygrometer
Hide thumbs Also See for Moisture monitor series 3:
Table of Contents

Advertisement

Entering Oxygen Curve
Data
Advanced Programming
O2 Curve Menu
[S/N]
CURVE
BkGd
Probe Serial #
O2: 123456
O2 Curve Menu
S/N
[CURVE]
BkGd
Sel. O2 Curve Pts#
[ZERO]
SPAN
#1
O(uA)
O(ppm)
0.721
0.0000
#1
O(uA)
O(ppm)
0.721
0.5000
[NO]
Press the
key to return to the Select O2 Curve Points menu. Use
the arrow keys to select SPAN and enter the span microamp (IO1),
and ppmv or % (IO1) values.
You may now do one of the following:
Enter data for a different channel or type of sensor by pressing the
[NO]
key until you reach the Programming Menu, and then refer to
Entering Calibration Data for New Probes/Sensors on page 3-39.
Refer to another section and perform a different procedure. Refer
to Figure 3-2 on page 3-48 and Figure 3-3 on page 3-49 to navigate
through the user program.
[RUN]
Exit by pressing the
IMPORTANT:
You must configure a channel AND enter the
necessary calibration data in order for a sensor to
work properly.
1
Use the arrow keys to move the
brackets to S/N and press
1
Enter the serial number and press
[YES]
. Refer to the Calibration
Data sheet for the serial number.
[NO]
Press
to return to the Curve
Menu.
1
Use the arrow keys to move the
brackets to CURVE and press
[YES]
.
Use the arrow keys to move the
brackets to ZERO and press
[YES].
Enter the zero microamp value
(IO0) that is shown on the
Calibration Data Sheet. Press
[YES]
and press the right arrow
key.
Enter the zero ppmv (or %) value
[YES]
(OX0) and press
key.
March 2008
[YES]
.
.
3-43