Entering Calibration Data For New Probes/Sensors - GE Moisture monitor series 3 Programming Manual

Panametrics hygrometer
Hide thumbs Also See for Moisture monitor series 3:
Table of Contents

Advertisement

Entering Calibration
Data for New Probes/
Sensors
Advanced Programming
The MMS-3 needs calibration data for each moisture, oxygen and
pressure sensor, as well as each auxiliary input. GE Sensing supplies
each probe with a Calibration Data Sheet, which is a list of 1 to 16
data points the electronics needs to interpret raw data from the probe.
Calibration Data Sheets are packed with each sensor.
Note: If you want to enter a constant value for moisture,
temperature, pressure, or multiply the ppm
constant multiplier, refer to Entering System Constants on
page 3-26. It is not necessary to enter the calibration data if
you are using a constant.
Note: If you have already entered the System Menu, and selected a
channel and measurement mode, proceed to the CURVES
selection step.
[PROG]
Press the
key to enter the user program..
Enter Passcode: XXXX
Be sure the number displayed in the upper right-hand corner of the
screen is the channel you want to program. If not, press the
key to select the desired channel.
Programming Menu
[SYSTEM]
AUTOCAL
.
Measurement Mode
O
[H]
T
P
AUX1
value by a
v
.
Enter the passcode
1
Use the arrow keys to move the
brackets to SYSTEM and press
[YES].
1
You must now select the type of
measurement the probe is
capable of taking. Use the arrow
keys to move to the desired
measurement mode and press
[YES].
Refer to Table 3-12 on
page 3-40 for a list of available
measurement modes.
March 2008
[CHAN]
3-39