Multiple Input Instructions - GE Moisture monitor series 3 Programming Manual

Panametrics hygrometer
Hide thumbs Also See for Moisture monitor series 3:
Table of Contents

Advertisement

March 2008
Reconfiguring a Channel
for a New Sensor (cont.)

Multiple Input Instructions

3-36
System Menu
[CONFIG]
REF
Sel. HYGRO sensor
---
TF
[Mxx]
Note: The word HYGRO will be replaced to reflect the measurement
mode you previously selected.
Table 3-10 on page 3-37 shows a list of measurement modes that
require multiple inputs and what you need to display them.
Example 1:
If you are using the TF Probe on channel 1, you must configure that
channel for moisture, temperature and pressure. To do this, you must
select TF under the CONFIG option for the H, T and P measurement
modes.
Example 2:
If you are using an M Series Probe to measure moisture and a
constant pressure on channel 1, you must configure that channel for
both moisture and a constant pressure. To do this, you must select
Mxx under the CONFIG option for the H measurement mode. You
will also have to select Kp under the CONFIG option for the P
measurement mode.
To select another measurement mode, press the [NO] key until you
return to Measurement Mode and repeat the following steps to
complete configuration. See Figure 3-2 on page 3-48 and Figure 3-3
on page 3-49 for menu maps.
When activating pressure sensors, the unit will prompt you to select
which input (Aux 1 or Aux 2) the pressure sensor is connected to.
After selecting the input, the display will prompt you to enter the
probe signal in either current (I) or volts (V).
1
Use the arrow keys to move the
brackets to CONFIG and press
[YES].
1
Use the arrow keys to move the
brackets to the desired selection
[YES].
and press
Refer to Table 3-
11 on page 3-38 for a list of
available choices for each
sensor. If you want to use a
constant, select Kh for moisture,
Kt for temperature, or Kp for
pressure.