Reviewing the calibration

NOTE:
Monitor windows, refer to the online Operator's Guide.
Use the View Calibration Curve window to review the calibration of photometric
methods.
Use the ISE Monitor window to review the calibration of electrolyte methods.
Calibration/RBL History Window
Use the Calibration/RBL History window to review the calibration history for all
methods that are run at the system and to create CSV files.
1. At the Menu Panel, select Calibration, then select Calibration/RBL History.
2. Select the calibration data to display by selecting Latest or Select:
• Latest–Displays the most recent calibration data for each method on the system.
• Select–Activates the Test Name box to allow a more detailed test selection.
3. If the Select option is chosen, then select one of the 3 Test Name options:
• Photometric–Displays all calibration history data for all photometric methods.
• ISE Tests–Displays all calibration history data for all ISE methods.
• Other–Activates the test list box to allow specific test selection (highlight each
4. Specify a date range, to display data for calibrations performed on specific days for
the tests selected:
a. Type the Start Date (yyyymmdd) of the range (if other than the default, which is
b. Type the End Date (yyyymmdd) of the range (if other than the default, which is
c. Select Search to display the calibration data for the tests selected for the
NOTE:
calibration data table.
d. To clear the data for the method selected, select Clear.
e. When prompted, select Yes.
5. To reset the calibration select the Reset check box.
NOTE:
column an save your changes, those changes are retained when you exit out and back
into the window.
6. Select Print, to print the data displayed.
7. To create a CSV file of the displayed data for archiving, or for exporting the data to
another application, do the following:
a. First, select the data you want to save.
96
For detailed information on how to use the View Calibration Curve and ISE
test to display).
the current date).
the current date).
specified date range.
Select and drag the column heading borders to resize the columns in the
You cannot cancel the reset once you select Save. If you resize the reset
ADVIA 2400 Operator's Guide