Washing The Ise Electrodes Lines - Siemens ADVIA 2400 Operator's Manual

Chemistry system
Hide thumbs Also See for ADVIA 2400:
Table of Contents

Advertisement

Figure 5-22. Locating pin
CAUTION
Do not touch the glass portion of the lamp. Touching the lamp could damage it. If the
lamp is dirty, clean it using lint-free toweling moistened with ethanol.
6. Install the lamp screws.
7. Install the lead wires and fasten the knobs.
8. Replace the access panel.
9. Return the analyzer to the READY mode, if necessary.
10. Wait 40 minutes for the lamp to stabilize.
NOTE:
11. Then, check the lamp energy.
12. Register the data and verify that attenuations register at 100%, voltages are between
5–9, and scatter is no greater than ±40.
13. Perform the cell blank measurement test.
NOTE:
replaced.
14. Run QC controls to verify that all assays are within the laboratory's established
control ranges.

Washing the ISE electrodes lines

Materials required:
• Dummy electrode with o-ring and cap
(REF 05938765, PN 073-0342-01)
• ISE detergent solution
(REF 01307361, B01-4174-01)
Time: 25 minutes
Analyzer mode: Manual operation
Maintenance
You can perform the daily shutdown 2 while you are waiting.
Siemens recommends the assays on the system be calibrated after the lamp is
BIOHAZARD
Wear personal protective
equipment.
Use universal precautions.