Every 3 Months Maintenance; Replacing The Lamp - Siemens ADVIA 2400 Operator's Manual

Chemistry system
Hide thumbs Also See for ADVIA 2400:
Table of Contents

Advertisement

4. Remove the bottle.
CAUTION
Make a note of the bottle position on the shelf, to avoid mixing up the fluid bottles.
5. Empty any remaining contents of the bottle.
6. Rinse the bottle with deionized water and drain well.
CAUTION
Ensure that the level-sensor connector does not get wet to avoid damaging it.
7. Refill the bottle with cuvette conditioner.
8. Return the bottle to the same position on the shelf in the cabinet.
9. Connect the cuvette-conditioner bottle level sensor connector, then push the
connector in and turn clockwise.
10. Insert the filter and hose, then fasten the cap.
NOTE:
11. Prime the fluid lines:
NOTE:
a. At the Operation Panel, select Prime.
b. At the Prime Settings dialog box, select Prime 2 and type 10 for the Number of
c. Select Execute.

Every 3 months maintenance

Replacing the lamp

Materials required:
• Halogen lamp, 12 V/50 W
(REF 02127928, PN 073-0099-01)
Time: 60 minutes
Analyzer mode: OFF
Maintenance
Make sure that the filter holder is located at the bottom of the bottle.
If you are cleaning other bottles, wait to perform this step for all fluid lines.
times in all fields.
BIOHAZARD
Wear personal protective
equipment.
Use universal precautions.
133