Siemens ADVIA 2400 Operator's Manual page 191

Chemistry system
Hide thumbs Also See for ADVIA 2400:
Table of Contents

Advertisement

Item
Minimum sampling volume
Special dilution
Dilution
Dilution tray (DTT)
Number of cuvettes
Maximum sample volume
Dilution cuvette dead volume
Dilution ratio
Reaction cuvette material
Reagent
Dispensing system
Trays
Container capacity
Refrigerator
Reagent volume/item
Dilution reagent volume
Reagent inventory
Barcode
Reaction
Reaction tray (RRV)
Number of cuvettes
RRV cuvette material
Reaction liquid volume
Stirring system
Reaction times
Reaction temperature
Reaction tank
Appendix
Description
2 µL (0.1 µL increments)
Diluted sample can be rediluted directly from tray.
Turntable system
120 (6 sets of 20 cuvettes)
300 µL
35 µL
From 1 to 1:5625 (1:75 x 1:75)
Plastic
2-reagent capability, 2-probe system
Two trays, each holding 50 containers
Multiple reagent pack loads – 5 reagents per method with automatic
rollover upon depletion
Reagents on each tray are cooled to between
6 °C and 14 °C
20 or 70 mL. (40 mL with adapter)
Recirculated liquid
10 to 100 µL (0.1 µL increments)
10 to 100 µL (0.1 µL increments)
Computed by the liquid-level sensing
Interleaved 2 of 5. JSCLA standard.
Turntable system
340 (20 sets of 17 cuvettes)
Plastic
60 to 180 mL
Rotation and reciprocation
Strong and weak stirring options are available.
Stirring immediately after additions of R1 and sample and R2
Mixer 1 is used for R1; Mixer 2 is used for R2.
3, 4, 5, 10 minutes
Extended reaction times: 15 & 21 minutes
37°C
Temperature regulation: ±0.1°C
Inert liquid circulation system
191