Daily maintenance

Inspecting and cleaning the probes

Materials required:
• Phillips screwdriver
• Alcohol prep pad or lint-free towels and
5% bleach solution
Time: 10 minutes
Analyzer mode: READY
1. Visually inspect each probe daily.
Figure 5-1 Location of probes
Perform the Shutdown Wash and Weekly Wash as scheduled, to prevent the probes
from clogging.
2. Clean each probe using one of the 3 methods described below:

Cleaning the probes using automatic advance probe motion

1. At the Menu Panel, select
2. At the Manual Operation window, double-select the
move as follows:
Probe
dilution probe
sample probe
Reagent probe 1
Reagent probe 2
Maintenance
Replace any clogged probes. See
Guide.
Clean all probes daily (proceed to step 2).
TIP
Code
3.DPPLR
16.SPPLR
37.RPPLR-1
49.RPPLR-2
BIOHAZARD
Wear personal protective
equipment.
Use universal precautions
Replacing the Probes
, then select
Maint
Manual Operation
.
1)
Dilution Probe (DPP)
2)
Sample Probe (SPP)
3)
Reagent Probe 1 (RPP1)
4)
Reagent Probe 2 (RPP2)
in the online Operator's
.
for the probe you want to
code
103