Cuvette Wash And Cuvette Conditioner Usage; Cleaning And Replenishing The Dilution Bottle - Siemens ADVIA 2400 Operator's Manual

Chemistry system
Hide thumbs Also See for ADVIA 2400:
Table of Contents

Advertisement

Cuvette wash and cuvette conditioner usage

CTT and RTT
10% Cuvette
Wash
System solutions
Cuvette Wash
Cuvette
Conditioner
NOTE:
vary slightly from the volumes provided in the tables above. This is normal behavior.

Cleaning and replenishing the dilution bottle

Materials required:
• Deionized Water
• Physiological saline (0.9% NaCl)
Time: 10 minutes
Analyzer mode: READY
NOTE:
once a month.
Figure 5-17. Isotonic Saline diluent bottle
128
Solution
Approximate
Volume Used
During Wash 2
DPP = 3.6 mL
RPP1 = 29 mL
RPP2 = 29 mL
Solution
Dilution by
system
1:10 with
water
1:40 with
water
The total volume of cuvette wash and cuvette conditioner used by the system can
The dilution bottle may be cleaned when it is refilled, but must be cleaned at least
Number of
Aspirations
DPP = 0.36 mL
1220 (approx)
RPP1 = 2.9 mL
RPP2 = 2.9 mL
Volume
Volume of
dispensed
undiluted
per WUD
solution
cycle
used per
WUD cycle
600 µL
60 µL
600 µL
15 µL
BIOHAZARD
Wear personal protective
equipment.
Use universal precautions.
1)
ISE buffer bottle
2)
RRV (Reaction) bath oil
3) Isotonic saline diluent bottle
4) Cuvette detergent bottle
5) Cell conditioner bottle
ADVIA 2400 Operator's Guide
Total number of
WUD cycles for
Wash2
Total volume
used for
Wash2
undiluted
44 ml
11 ml