Analytical Parameters (Chemistry) Window - Siemens ADVIA 2400 Operator's Manual

Chemistry system
Hide thumbs Also See for ADVIA 2400:
Table of Contents

Advertisement

The settings below are related to serum information items (serum indices), which are
involved in creating workorders. If the serum indices for each sample are not selected in
the Order Entry window, select one of these settings:
Request. item range
Compulsory item analysis
Not handled
Set the system monitor
Set system parameters
Basic system composition
These settings allow you to customize general components of the system:
ISE
Sample bar code
Concentrated waste
Basic system operation
These settings are used for retesting of samples
Auto. retest
Manual retest
Sample containers
The settings below specify the sample containers used by the system. You can specify up
to 9 container types.
CAUTION
Normal operation and the volume of sample aspirated depend on the specifications
you enter here. Incorrect settings may damage the probes and nozzles.
Type
Container diameters
LLS sensitivity
Reagent bottles
The settings below specify the reagent bottles used by the system. These settings are used
in the Reagent Inventory, Reagent Container Set, and Lamp Energy Monitor windows.
CAUTION
Incorrect settings may result in abnormal data.
Type
Bottle section

Analytical Parameters (Chemistry) window*

Use this window to set up chemistry methods.
Siemens methods are predefined when the system is received. For predefined methods,
only expected values, rerun conditions, and correction formulas should be modified.
You can define new methods. The system can store up to 200 chemistry methods in this
window (including predefined methods).
System Overview
Sample delivery
On-line
Reagent bar code
Container name
Container heights
Liq volume judge
Container name
LLS sensitivity
55