Priming the system

1. At the Menu Panel, select
refer to Using the User Maintenance window.).
2. At the User Maintenance window, in the Position Probes for Routine Cleaning area,
select
3. Ensure that the probe is perpendicular to the arm and centered over the cuvette.
4. If not centered, call your local technical support provider or distributor.
5. At the Operation Panel, select
wash cups).
6. At the Operation Panel, select
water flow through the probe.
NOTE:

Replenishing the RRV (reaction) bath oil bottle

Materials required:
• RRV (reaction) bath oil
(REF 09323099, PN B01-4180-01)
Time: 10 minutes
Analyzer mode: READY
Figure 5-33. RRV bath oil bottle
CAUTION
Do not attempt to clean the RRV bath oil bottle (
water do not mix.
1. Unscrew the filter cap (
tube with the filter.
2. Disconnect the RRV bath oil bottle level sensor connector (
connector counter-clockwise and pull it out.
150
Maint.
, then select
when prompted.
Start
Yes
Make sure that no water is leaking from the joint connector (4).
) at the front of the RRV bath oil bottle (
3
, then select
User maint
to return the probes back to home (over the
Initialize
,
, and then
Prime
PRIME 2
BIOHAZARD
Wear personal protective
equipment.
Use universal precautions.
1) ISE buffer bottle
2) RRV (Reaction) bath oil bottle
3) Isotonic saline diluent bottle
4) Cuvette detergent bottle
5) Cell conditioner bottle
) with water; RRV bath oil and
1
. (For additional details,
to ensure proper
Execute
), then pull up the
1
), then turn the
2
ADVIA 2400 Operator's Guide