ISE Monitor window

Perform any of the following tasks at this window:
monitor and verify calibration data as a function of time, using history graphs
monitor serum or urine calibrations
monitor calibration trace data
display data from past calibrations
*monitor selectivity check data
print the results
delete calibration trace data
change y-axis scale
transfer calibration data to the ISE
* The selectivity check must be performed under the supervision of authorized Siemens
service personnel. Please call your local technical support provider or distributor.

Manual Operation window

Perform any of the following tasks at this window:
operating a unit
activating or deactivate a unit
checking unit position values
BIOHAZARD
Wear personal protective equipment. Use universal precautions
Manual operation of units may be necessary when performing maintenance and
troubleshooting tasks.
The window contains a graphical image of the analyzer which depicts the operation you
are performing.
IMPORTANT
The operating mode must be READY or WAIT for you to run manual operations.
Also, if you cannot perform an operation due to system restrictions, a buzzer goes off and
a message displays at the Alarm message box (in the Operation Panel). Select
the Operation Panel for more information and instructions regarding the buzzer.

On-Line Monitor window

Use this window to monitor the exchange of data between the system and the host
computer. The monitor displays the data being exchanged and the control signal codes
that ensure accurate transmission of information.
You can view up to 300 data items in the window.