4.2 Three-Point Calibration (Linearity adjustment)

For the GP-X series, linearity adjustment in respect to the distance is made
before shipment from the factory with a set of a sensor head and controller.
If adjustment is made for the material to be detected and operating environ-
ment, linearity with increased accuracy is obtained.
<Setting 1 mode>
<Calibration>
Ramco Innovations www.ramcoi.com
Setting example (In the case of GP-XC5S)
Press
to select "cALIb".
If the calibration position is unsuitable, an error is displayed. Press the
To cancel calibration in the middle of the procedure, press the
key.
Save common calibration data to
four memories.
Make the object come in contact
with the sensor head.
Press the
key at the center
position (half the scale) of the
detection range.
Press the
maximum detection distance
(full-scale) of the detection
range.
To set adjustment, press the
Press the
key to return.
key to cancel and enter again.
Phone 800-280-6933 nsales@ramcoi.com
Object
Sensor head
Half the scale
(0.5mm)
key at the
Full-scale
(1mm)
key.
71