Commissioning
5.2 Local display
Hazardous contact voltage
Risk of injury through hazardous contact voltage when the device is open or not completely
closed.
The degree of protection specified on the nameplate or in Technical specifications (Page 97)
is no longer guaranteed if the device is open or not properly closed.
• Make sure that the device is securely closed.
Risk of explosion when media above 100 °C flows through the process flange
Explosion protection is no longer guaranteed and the approval is nullified.
It is prohibited for media above 100 °C to continually flow through the process flange.
Loss of explosion protection
Risk of explosion in hazardous areas if the device is open or not properly closed.
• Close the device as described in Connecting (Page 35).
Hot surfaces
Risk of burns resulting from hot surfaces.
• Take corresponding protective measures, for example by wearing protective gloves.
Note
Verify proper operation and configuration of all analog inputs
Incorrect configuration can have a negative impact on flow compensation and produce errors in
flow rate.
5.2

Local display

The device is commissioned/operated with the buttons on the local display.
The buttons are actuated by touching the glass panel on the appropriate button. The
graphical display above the buttons gives a menu-guided operation of the individual device
86
WARNING
WARNING
WARNING
WARNING
Compact Operating Instructions, 12/2023, A5E38755607-AF
SITRANS FS130 & FS230