7.8 Construction
Table 7-17
 
C
i
L
i
Table 7-18
 
U
o
I
o
P
o
C
o
L
o
L
/R
o
Note
External capacitance and inductance
The C
capacitance and inductance. No further reduction of external capacitance and inductance is
required.
7.8

Construction

Table 7-19
Description
Measurement of process medium
104
Siemens SSL cable capacity
SSL cables
264.6 pF/m
1.12 µH/m
Output parameters
RTD 5, RTD 6 (any combination of con‐
nections)
4.52 V DC
13.8 mA
26.6 mW
 
Group IIC
4.28 μF
Groups IIB, III
139.5 μF
Group IIA
994 μF
 
Group IIC
1.87 mH
Groups IIB, III
7.47 mH
Group IIA
14.9 mH
 
o
Group IIC
NA
Groups IIB, III
NA
Group IIA
NA
and L
values already take into consideration the combined effects of external
o
o
Designated use
Specification
Fluid Group 1 (suitable for dangerous fluids)
Aggregate state: Paste/light slurry, liquid and gas
Compact Operating Instructions, 12/2023, A5E38755607-AF
Sensor outputs, each
(up to 8)
17.43 V DC
452 mA
991 mW
 
169.5 nF
985 nF
1000 nF
 
67 μH
348 μH
697 μH
 
18.0 μH/Ω
72.2 μH/Ω
144 μH/Ω
SITRANS FS130 & FS230