Connecting The Ssl Cable (Wall Mount Enclosure) - Siemens SITRANS FS130 Compact Operating Instructions

Ultrasonic flowmeters
Table of Contents

Advertisement

Connecting the flow sensor cables
When connecting the sensors it is required for the system to function that the sensor cables are
connected correctly. Both sensor cables for each path must be connected to the same port
numbers (1A and 1B or 2A and 2B) on the transmitter.
The sensor cables are delivered with a premounted F-connector at the transmitter end.
1. Remove the blind plug from the wall mount enclosure.
2. Slide the cable gland back onto the cable and allow access for the F-connector tool.
3. Place F-connector tool around cable and slide up to engage F-connector nut.
4. Push cable through the gland opening. Ensure that the center lead is aligned with connector
port in the transmitter.
5. Tighten the F-connector using the tool until you meet a mechanical stop.
6. Remove the F-connector tool.
7. Mount and tighten the cable gland.
Repeat these steps for each sensor cable.
4.7.3

Connecting the SSL cable (wall mount enclosure)

M12 connector version
The SSL cable is provided with M12 style stainless steel plugs.
SITRANS FS130 & FS230
Compact Operating Instructions, 12/2023, A5E38755607-AF
Path 1 - Upstream sensor cable (1A)
Path 1 - Downstream sensor cable (1B)
4.7 Connecting wall mount enclosure
Path 2 - Upstream (2A)
Path 2 - Downstream (2B)
Connecting
69