HP 3457A Service Manual page 86

Hide thumbs Also See for 3457A:
Table of Contents

Advertisement

3457A
Operational
Verification
Tests
4-Wire Ohms Operational Verification Tests Cont'd
34570
3457A
| 3457A
90 day limits
1 year Limits
Input
Range
| Set Up
High
Low
High
Low
3 Kohm
3 Kohm|
OHMF
3.000112 Kohm
2.999888 Kohm
3.000157 Kohm
2.999843 Kohm
30 Kohm|
30 Kohm|
OHMF
30.00112 Kohm
29.99888 Kohm
30.00157 Kohm
29.99844 Kohm
300 Kohm|
300 Kohm|
OHMF
300.0128 Kohm
299.9872 Kohm
300.0158 Kohm
299.9842 Kohm
Table 4-28. 4-Wire Ohms Gain Test Limits
Frequency Counter Operational Verification Tests
FREQUENCY
SOURCE
Figure 4-19, Frequency Counter Operational Verification Test Connections
4-88. Frequency Counter - Accuracy Test
4-89. Equipment
Required.
A Frequency Source capable of providing a 20 Hz (+ 01%) and a | MHz
(+ 003%) sine-wave signal is required for this procedure.
1, Set the HP 3457A to the Frequency
Function (FREQ).
2. Set the Frequency Standard for a | volt, 20 Hz sine-wave output signal and connect it to the HI
and LO input terminals of the HP 3457A.
3. Use the Test Record to record the frequency readings at 20 Hz and | MHz.
4. If either of the readings are beyond
the limits specified, the instrument
should
be calibrated.
Refer to Section V of this manual for calibration procedures.